Recall Z-0300-2022 ยท OWB Interventional Fluoroscopic X-Ray System
Siemens Medical Solutions USA, Inc
Artis Q floor with software VD12- angiography systems developed for single and biplane diagnostic imaging and interventional procedure Model Number: 10848280
| Date | Oct 20, 2021 |
| Status | Terminated |
| Reason | If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure |
| Root cause | Software design |
| Action | Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice dated 10/202/1 to customers via AX070/21/S, Siemens will update the system software via Update Instruction AX069/21/S. Letter states reason for recall, health risk and action to take: To avoid the problem, it is recommended to position the SID first and then release x-ray. Siemens will correct the software error (software VD12) via Update Instruction AX069/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organizatio |
| Quantity | 8 units |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181407 | Artis zee/zeego & Artis Q/Q.zen Siemens Medical Solution USA, Inc., Aug 15, 2018 |