Recall Z-0293-2022 ยท EZW Stimulator, Electrical, Implantable, For Incontinence

Medtronic Neuromodulation

Medtronic Wireless Recharger Model WR9220 included in the RS5200 Recharger Kit used by the InterStim Micro (Model 97810) implantable neurostimulator.

DateNov 4, 2021
StatusOpen, Classified
ReasonThe wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
Root causeOther
ActionThe recalling firm issued undated letters on 11/4/2021 via USPS Priority Shipping. The recalling firm has three different letters addressed to the Healthcare Professional depending upon the type of device the wireless recharger is used with (InterStim Micro, product number RS5200; Activa RC, product number RS6200; or RestoreUltra & RestoreSensor, Product number RS7200, which is currently not distributed in the U.S. at this time). The letter gives an issue overview and provides requested actions. Those actions include reinforcing the instructions in the enclosed Patient Quick User Guide wit
Quantity10,081 devices
DistributionUS Nationwide, including PR. There was government distribution but no military distribution was identified.