ClearanceBrief
Recall Z-0290-2022 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

Medtronic Vascular, Inc.

Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE

DateOct 15, 2021
StatusOpen, Classified
ReasonDuring stent graft deployment, the spindle may detach from the distal end of the spindle hypotube
Root causeUnder Investigation by firm
ActionOn or about 10/15/2021, the firm initiated a global verbal notification to customers informing them that is the potential for a specific subset of 18Fr Stent Graft Systems to have the potential for the spindle to detach from the spindle hypotube during deployment. On 10/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-day delivery. Customers are instructed to take the following actions: 1) Return all unused affected devices to the Recalling Firm via their sales representative or calling customer service at 1-888-283-7868 or email rs.aorticorders@medtroni
Quantity2 devices
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of CA, FL, GA, MN, MO, NC, NJ, NM, NV, NY, PA, SC, TN, TX, VA, and WV. The countries of Austria, Belgium, Bulgaria, China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hungary, Ireland, Italy, Japan, Jordan, Lithuania,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.