ClearanceBrief
Recall Z-0312-2022 ยท PEN Gram-Negative Bacteria And Associated Resistance Markers

BioFire Diagnostics, LLC

BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog No./Description: 442023/BD BACTEC Plus Aerobic medium in plastic vials 442021/BD BACTEC Lytic Anaerobic medium in plastic vials 442192//BD BACTEC Plus Aerobic medium in glass vials 442022/BD BACTEC Plus Anaerobic medium in plastic vials 442020/BD BACTEC Peds Plus mediums in plasti

DateOct 11, 2021
StatusTerminated
ReasonDue to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
Root causeMaterial/Component Contamination
ActionOn or about 10/10/2021, the Recalling Firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to U.S. customer via overnight FedEx to inform them that there is an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results when the blood culture identification panels are used with some types of blood culture vials. The increased risk are potential due to an increased level of nucleic acid from non-viable Acinetobacter calcoaceticus- baumannii complex targets in the blood culture vials. Affected Blood Culture Media vials are: 442023 - BD BACTEC Plus Ae
QuantityTotal 14,140 Kits [11,155 kits (U.S.) and 2,985 kits (O.U.S.)]
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentin
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193519BioFire Blood Culture Identification 2 (BCID2) Panel
Biofire Diagnostics, LLC, Mar 18, 2020