Recall Z-0244-2022 ยท LXH Orthopedic Manual Surgical Instrument

Laboratoires Bodycad, Inc.

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.

DateJul 2, 2021
StatusTerminated
ReasonThere is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Root causeNonconforming Material/Component
ActionThe recalling firm issued an Advisory Notice Information dated 2020-06-30 on 7/2/2021 via email which explained the reason for recall and that they would be contacted as soon as possible to proceed with the exchange. On 9/24/2021, a Returned Goods Authorization Form was sent to the consignees so the adapter could be exchanged.
Quantity10 in the U.S.
DistributionDistribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.