Recall Z-0267-2022 ยท ECX Cylinder, Compressed Gas, And Valve

Western/Scott Fetzer Company

Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840

DateOct 16, 2021
StatusOpen, Classified
ReasonVIPR (Valve Integrated Pressure Regulator) - battery rupture, causing the plastic VIPR shroud to break apart with emission of fumes and smoke
Root causeDevice Design
ActionWestern / Scott Fetzer Co. issued Urgent Medical device Recall letter, dated 10/15/21 distributed via USPS to consignees on 10/16/21, and emailed to the same consignees on 10/18/21. Letter states reason for recall , health risk and action to take: Fill facility employees are asked to verify the functionality of each unit through the observation of information on the digital display. Note that this direction and expectation is already clearly indicated in the product Operating Instructions and Inspection Guidelines. If the digital display shows the expected information as depicted in the Operat
Quantity11341 units
DistributionCT, MD, MT, OH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.