Recall Z-0226-2022 ยท MUQ Glue, Surgical, Arteries

Siemens Medical Solutions USA, Inc

Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo Model 11007641 (3)Sensis Vibe Combo Model 11007642 Intended Use: A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies

DateOct 5, 2021
StatusOpen, Classified
ReasonSystem operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs
Root causeSoftware design
ActionSiemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to customers via AX067/21/S dated 10/5/21. Letter states reason for recall,health risk and action to take: The software in the affected systems will be updated to display a notification message. Once the system has reached a runtime of more than 7 days, a notification message will be displayed after closing a study to remind the operator to reboot the system as it is running longer than advised. In addition, the system operating manual will include a warning message that the Sensis system can have performance is
Quantity403 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150493Sensis
Siemens Medi Cal Solutions, Inc., Jun 30, 2015