Recall Z-0367-2022 ยท FWM Prosthesis, Breast, Inflatable, Internal, Saline

Mentor Texas, LP.

MENTOR Saline-Filled Breast Implants are constructed from room temperature vulcanized silicone elastomer, made of polydimethylsiloxane. Surgeons inflate the silicone elastomer shell to the desired size with sterile isotonic saline before implantation. The implants are available with SILTEX Textured or smooth surface shells. Each implant is supplied sterile with a disposable fill tube and reflux va

DateSep 30, 2021
StatusOpen, Classified
ReasonDue to a manufacturing issue, Implant may have a potentially weakened area on the base patch portion of the implant which may cause leakage/deflation of the affected device.
Root causeProcess control
ActionOn 10/11/2021, the Firm sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL)" Notfication via UPS informing customers that breast implants which have expiration dates from 01/01/2025 to 09/30/2021, due to a manufacturing issue may have a potentially weakened area on the base patch portion of the shell. As a result of this issue, these implants may prematurely deflate for which replacement surgery may be needed. Customers are instructed to: 1. Examine inventory immediately to determine if they have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject
Quantity28,348 implants
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.