Recall Z-0193-2022 ยท CDT Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides

Siemens Healthcare Diagnostics, Inc.

Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1 and lower. SMN 10703038

DateOct 6, 2021
StatusOpen, Classified
ReasonThe Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical interpretations and clinical decisions.
Root causeSoftware design
ActionOn October 6, 2021, the firm notified affected customers via an "Urgent Medical Device Correction" letter. Customers were informed of the product issue and provided instructions on actions to be taken by laboratories, which prevent the issue from happening. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. A new version of the Firmware for the Aptio Automation Interface Module to the ADVIA Centaur XP/XPT will correct the issue. Your service provider will contact you to schedule the f
Quantity1142
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043546DIMENSION RXL MAX WITH STREAMLAB ANALYTICAL WORKCELL AND SAMPLE TRANSFER MODULE
Dade Behring, Inc., Jan 18, 2005