Recall Z-0182-2022 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

DateOct 13, 2021
StatusOpen, Classified
ReasonUnder certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.
Root causeSoftware design
ActionAn Urgent Important Field Safety Notification letter (FAC-NU-0007) was issued to consignees to notify of the issue, clinical impact and recommend the following: 1) Recommended User Action in the unique scenario, which is only applicable for the Venezia" applicator: Avoid use of the combination of the two Venezia" applicator Anchor Points O1-Center and O2-Center by disabling one of them in the Applicator Library Manager. 2)Recommended User Action in the exceptional situation that may also occur for other Applicator Models: Avoid incorrect placement of the Anchor Points, especially exchanging
Quantity1072 devices
DistributionDistribution throughout USA OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, German
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132816ONCENTRA BRACHY
Nucletron B.V., Jun 13, 2014