Recall Z-0213-2022 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America Llc

Philips Azurion Interventional Fluoroscopic X-Ray System, Software version 2.1.x, Model numbers 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222, 722223, 722224, 722225, 722226, 722227, 722228. (Azurion 3 M12, Azurion 3 M15, Azurion 5 M12, Azurion 5 M20, Azurion 7 B12, Azurion 7 B20, Azurion 7 M12, Azurion 7 M20)

DateSep 24, 2021
StatusTerminated
ReasonWhen the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Root causeSoftware design
ActionOn September 24, 2021, the firm sent "Urgent Medical Device Correction" recall letters to affected customers (medical facilities). Customers were informed that a software defect causes the possibility for table lock-up and unavailable X-ray imaging when the APC (accept) button and the Float Tabletop (panning) button are pressed or released at the same time. If the problem occurs, there will be a delay in procedure. ACTIONS FOR THE CUSTOMER: - Do not press or release both the APC (accept) button and the Float Tabletop (panning) button at the same time during the procedure. - If the table lo
Quantity428 (US); 581 (OUS)
DistributionDomestic distribution nationwide. International distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018