Recall Z-0243-2022 ยท JSC Culture Media, General Nutrient Broth

bioMerieux, Inc.

API 50 CH, Model 50300

DateSep 30, 2021
StatusTerminated
ReasonThere is potential to provide an incorrect organism identification.
Root causeMaterial/Component Contamination
ActionNotification letters were issued to customers via FedEx Priority Overnight mail during the week of October 4, 2021. Customers are asked to do the following: -Discontinue use of and discard any remaining inventory of lot 1008679230. -For tests previously performed using API 50 CH Ref. 50300 lot# 1008679230, we are asking you to identify any possible incorrect identification test result due to a false positive reaction of test 34, analyze the related risks and to determine appropriate actions, if relevant. -Complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux
Quantity1382 kits
DistributionDistribution in WI, NY, and SD