Recall Z-0354-2022 ยท KNW Instrument, Biopsy
Argon Medical Devices, Inc
SuperCore Biopsy Instrument: REF/Product Description: 701114090/SUPERCORE BIOPSY INSTR 14GA X 9CM; 701114150/SUPERCORE BIOPSY INSTR 14GA X 15CM; 701116090/SUPERCORE BIOPSY INSTR 16GA X 9CM; 701118060/SUPERCORE BIOPSY INSTR 18GA X 6CM; 701118090/SUPERCORE BIOPSY INSTR 18GA X 9CM; 701118150/SUPERCORE BIOPSY INSTR 18GA X 15CM; 701118200/SUPERCORE BIOPSY INSTR 18GA X 20CM; 701120090/SUPERCORE BIOPSY
| Date | Oct 4, 2021 |
| Status | Open, Classified |
| Reason | Due to component (housing and plunger) detaching from the semi-automatic biopsy instruments. |
| Root cause | Device Design |
| Action | On 09/30/2021, the firm sent an "Urgent-Product Recall Notice" letter to customers via US Postal Service Registered Mail informing them that the firm has received complaints regarding the semi-automatic Biopsy instrument coming apart during shipping or prior to use. According to the firm's investigation, the plastic housing and plunger can be separated more easily than normal for the lots manufactured during this time frame. Customers are instructed to return all affected product to the Recalling Firm's Athens TX facility using RGA#26683, attention Andrea Wieczor. Mailing address is: RGA# 2 |
| Quantity | 79,910 devices |
| Distribution | Worldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of ALB, ARE, ARG, AUS, AUT, BEL, BGD, BRA, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K974814 | MANAN(TM) SUPER-CORE BIOPSY NEEDLES Medical Device Technologies, Inc., Jan 9, 1998 |