Recall Z-0077-2022 ยท PRL Iliac Covered Stent, Arterial

W L Gore & Associates, Inc.

Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diameter 16 mm, UDI:

DateSep 7, 2021
StatusOpen, Classified
ReasonIncorrect labeling, specifically, expandable endoprosthesis labeled as 8 mm x 59mm x 135 cm, however may contain a 9mm x 29 mm x 135 cm device.
Root causeComponent change control
ActionOn September 9, 2021, Gore issued a "Urgent Field Safety Notice" to affected consignees via E-Mail. In addition to informing consignees about the recalled product, Gore asked consignees to take the following actions: 1. Identify and return the device listed in the CHANNEL PARTNER RESPONSE FORM if unused. 2. Please complete and sign the enclosed CHANNEL PARTNER RESPONSE FORM and return to FieldActionTeam@wlgore.com within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. 4. If a listed device has been implanted, the
Quantity2 units
DistributionInternational distribution to the countries of Italy & Lebanon.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.