ClearanceBrief
Recall Z-0264-2022 ยท IOR Wheelchair, Mechanical

TiSport, Llc

Permobil TiLite, Models: Aero Z, and ZRA

DateSep 24, 2021
StatusOpen, Classified
ReasonSome affected wheelchairs, with transit tie downs, that are used as a passenger seat in a motor vehicle may not be configured in the right orientation to perform as intended if involved in a motor vehicle accident. Some affected wheelchairs can no longer be used as a passenger seat until transit tie downs are replaced with redesigned transit tie downs, verified, and released to the market.
Root causeComponent change control
ActionOn 09/24/21, distribution of correction notices started, via mail, to distributors. Distributors were advised the following: For AERO Z wheelchairs, the transit tie downs should have an inward orientation with transit bracket in rear most position. The wheelchair provider was provided with clear and thorough work instructions on how to inspect and, when necessary, correctly configure the transit tie downs. Stop using TiLite ZRA with transit tie downs as a passenger seat in a motor vehicle. Remove transit tie downs on TiLite ZRA wheelchairs. The recalling firm, if requested, will supply at no c
Quantity5373
DistributionUS Nationwide Distribution: MD, PA, VA, IN, NY, MO, AR, NE, NJ, WA, NM, IL, CA, OH, KY, LA, OK, WI, TX, NC, SD, MN, MA, OR; OUS (Foreign): AUSTRALIA, CANADA, FRANCE, GERMANY, IRELAND, ITALY, UNITED KINGDOM, URUGUAY, NEW ZEALAND, IRELAND, SWEDEN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990358CROSS SPORT, RIGID WHEELCHAIR
Tisport, Feb 26, 1999