Recall Z-0107-2022 ยท FRN Pump, Infusion

Smiths Medical ASD Inc.

CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L

DateOct 19, 2021 (initiated Mar 24, 2020)
StatusTerminated
ReasonSpecific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.
Root causeComponent change control
ActionThe firm notified its consignees by email beginning 03/24/2020. The notice explained the issue, risk to health, and requested a response. Distributors/pharmacy suppliers were directed to notify their customers. The firm is seeking return of the units for correction.
Quantity642 units
DistributionDistribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE