Recall Z-0087-2022 ยท CEM Electrode, Ion Specific, Potassium

Inpeco S.A.

FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2). (Different brand names of same product.)

DateSep 10, 2021
StatusOpen, Classified
ReasonThe Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.
Root causeSoftware design
ActionOn September 10, 2021, the firm distributed "Urgent IVD Medical Device Correction" letters to affected consignees. Distributors were instructed to forward the recall communication to downstream customers. Customers were informed that the potential risk associated with the issue is related to an incorrect patient result that may lead to incorrect patient treatment. Users were provided with detailed instructions on precautions to avoid the occurrence of the described issue, including a daily visual check and what to do in case of error 3160 or 0D60. Contact your local technical support provid
Quantity262 total, 52 US
DistributionDomestic distribution nationwide . Foreign distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121012FLEXLAB 3.6, ACCELERATOR A3600
Inpeco S.P.A., Aug 31, 2012