Recall Z-0215-2022 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

Hemodialysis Delivery System, Software Version 2.x.

DateSep 29, 2021
StatusOpen, Classified
ReasonIf the operator initiates therapy with a saved prescription profile and makes a change to the prescription after a disposable filter change using the Same Patient button, the system may suggest values from the original prescription profile, rather than the current prescription.
Root causeSoftware design
ActionAn Urgent Medical Device Correction communication was sent to affected customers via U.S.P.S., first class mail, on 09/29/2021. Baxter advises the following: 1. Operators may continue to safely use the PrisMax System with software version 2.x until the system is upgraded to version 3.x. To ensure patient safety, operators must follow the Operator s Manual and confirm all prescription fields are accurate prior to starting treatment, or when making prescription changes during therapy. The instructions can be found in the PrisMax Operating Instructions section of the Operator s Manual. Addition
Quantity4248
DistributionDistributed nationwide to AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC,SD, TN, TX, UT, VA, WA, WI and internationally to Mexico, Brazil, Canada, Australia, Singapore, Hong Kong, Vietnam, Austria, Belgium, Denmark, Finl
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K190910PrisMax System Version 2
Baxter Healthcare Corp, Jul 25, 2019