ClearanceBrief
Recall Z-0157-2022 ยท HAW Neurological Stereotaxic Instrument

Medtronic Xomed, Inc.

Medtronic sterile single-use blades that are components of the StraightShot M Series Handpiece and Fusion ENT Navigation Software, and the Midas Rex Microsaw and Triton System with the following product numbers: (1) REF 1883080EM, Rotatable Fusion Blade, Quadcut 3.0 mm x 13 cm; (2) REF 1883480EM, Rotatable Fusion Blade, Quadcut, 3.4 mm x 13 cm; (3) REF 1884006EM, Rotatable Fusion Blade, RAD 40; (

DateOct 21, 2021 (initiated Aug 31, 2021)
StatusOpen, Classified
ReasonThere was an increase in complaints related to tip breaks and wobble/vibration.
Root causeProcess control
ActionThe recalling firm issued letters dated 8/13/2021 on 8/31/2021 via USPS. The letter explained the reason for recall, risk to patient, and the actions to be taken. The consignee was requested to review their inventory of Medtronic blades against the list of affected products enclosed with the letter and to quarantine them. All unused, affected devices are to be returned to Medtronic by completing the Customer Confirmation form and returning it via FAX or email. Once completed, Customer Care will provide return instructions and issue a replacement for the returned product.
Quantity274,991 total devices
DistributionWorldwide distribution - US Nationwide, including PR. There was also government/military distribution and the countries of Australia, Austria, Bahrain, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Icela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130608QUADCUT
Medtronic Navigation, Inc., Aug 7, 2013