Recall Z-0024-2022 ยท HSN Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer

Exactech, Inc.

Vantage Total Ankle System including a. Vantage Fixed-Bearing Polyethylene Liner Component (Product Line: 350-21-XX, 350-22-XX); b. Vantage Mobile-Bearing Polyethylene Liner Component (Product Line: 350-41-XX, 350-42-XX). NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

DateOct 4, 2021 (initiated Aug 30, 2021)
StatusOpen, Classified
ReasonInserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Root causeProcess control
ActionExactech notified distributors and sales representatives on about 08/30/2021 via letter titled "URGENT MEDICAL DEVICE RECALL." Actions included removing all Knee and Ankle UHMWPE products labeled with an 8-year shelf life and not packaged in EVOH/Nylon bags, in a phased approach over 12 months. Phase 1: immediately return all knee and ankle UHMWPE devices labeled with an 8-year shelf life that will be 5 years old or older by 08/31/2022 not packaged in EVOH/Nylon bags. Phase 2: between 05/31/2022 to 08/31/2022, returning all remaining knee and ankle UHMWPE devices labeled with an 8-year she
Quantity2,217 devices
DistributionWorldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152217Exactech Vantage Total Ankle System
Exactech, Inc., Mar 10, 2016