ClearanceBrief
Recall Z-0020-2022 ยท HRY Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer

Exactech, Inc.

OPTETRAK Comprehensive Knee System including OPTETRAK All-polyethylene UNI Tibial Components, Product Lines, 252-12-XX, 252-13-XX, 252-22-XX, 252-23-XX. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

DateOct 4, 2021 (initiated Aug 30, 2021)
StatusOpen, Classified
ReasonInserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Root causeProcess control
ActionExactech notified distributors and sales representatives on about 08/30/2021 via letter titled "URGENT MEDICAL DEVICE RECALL." Actions included removing all Knee and Ankle UHMWPE products labeled with an 8-year shelf life and not packaged in EVOH/Nylon bags, in a phased approach over 12 months. Phase 1: immediately return all knee and ankle UHMWPE devices labeled with an 8-year shelf life that will be 5 years old or older by 08/31/2022 not packaged in EVOH/Nylon bags. Phase 2: between 05/31/2022 to 08/31/2022, returning all remaining knee and ankle UHMWPE devices labeled with an 8-year she
Quantity2,774 devices
DistributionWorldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040889EXACTECH OPTETRAK UNICONDYLAR KNEE
Exactech, Inc., Oct 1, 2004