Recall Z-0114-2022 ยท HTW Bit, Drill

Flower Orthopedics Corporation

Flower Lag Screw Kit 2.7 mm-intended to be used with 2.7mm lag screws and can be used with any plate that as part of the Flower Small and Medium Implant Set, Flower Ankle Plating Set , Flower Rear Foot/Calcaneus Plating Set , or Flower Upper Extremity Plating Set and uses the medium locking mechanism. Catalog Number: LSK 027

DateSep 1, 2021
StatusTerminated
ReasonDrill bits packaged in kits are missing the AO connect feature and may result in delay of procedure
Root causePackaging change control
ActionCoventus Orthopedic issued email communication on 9/01/21. Letter states reason for recall, health risk and action to take: Check your inventory for the product listed below. Next, please complete and return the attached Acknowledgement and Receipt Form by Friday, September 10, 2021. If you indicate on the form that you are in possession of affected product, Conventus Flower will ask that you return that affected kits. Conventus Flower will send replacement product as soon as we receive your completed Acknowledgement and Receipt Form. New DBK 020, DBK 024, DBK 027, or LSK 027 kits will be sh
Quantity1 unit
DistributionArizona California Colorado Florida Florida Illinois Indiana Nevada Ohio Texas Virginia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.