Recall Z-0141-2022 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

1.5T SIGNA HDxt MR System nuclear magnetic resonance imaging system

DateOct 16, 2021 (initiated Aug 26, 2021)
StatusTerminated
ReasonUnder certain conditions, some slices may be missing which can lead to a gap of anatomy in the 3D Volume images.
Root causeSoftware design
ActionGE issued an Urgent Medical Device Correction letter to customers with affected devices. This letter informs the customer of the issue and provides safety instructions for the continued use of the device. GE Healthcare will correct all affected devices at no cost to the customer. In the U.S., this letter will be sent by a traceable means. This letter will be sent to the following titles within the affected accounts: Director of Clinical/Radiology, Risk Manager/Hospital Administrator.
Quantity3 devices
DistributionDistribution in United States to AZ, CA, FL, GA, IL, LA, MA, MD, MN, NC, NH, NJ, NY, OH, OK, TX, VA, WA, and Puerto Rico OUS distribution to: Australia, Belgium, Canada, Chile, Colombia, France, Germany, Italy, Japan, Korea, Poland, Saudi Arabia, Spain, Sweden, Turkey, UAE, and UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052293GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM
Ge Medical Systems, LLC, Sep 21, 2005
K1216761.5T SIGNA HDXT 3.0T SIGNA HDXT SIGNA VIBRANT
Ge Medical Systems, LLC, Sep 20, 2012