Recall Z-0082-2022 ยท LKJ Antinuclear Antibody, Antigen, Control
Bio-Rad Laboratories, Inc.
BioPlex 2200 ANA Screen Control Set, REF: 663-1131 incompatible with ANA Screen Reagent Packs/ANA Kits, REF: 665-1150; and ANA Screen Calibrator Sets, REF: 663-1101
| Date | Nov 1, 2019 |
| Status | Terminated |
| Reason | A new lot of ANA screen control and any subsequent control lots are incompatible with older lot numbers of ANA calibrators and ANA reagent packs. |
| Root cause | Process control |
| Action | On 11/01/2019, ANA Screen Lot Matching-Compatibility Notice, were mailed to customers. Customers were advised of the following: If you have received the new control material, please ensure that your laboratory is using one of the newer lot numbers of BioPlex ANA calibrators and ANA reagents. Should you have any of the older BioPlex ANA reagent packs and calibrator sets listed below in ITALICS remaining in your inventory or if you have used these lots in combination with the new control, please contact your local Technical Support center. The following telephone number was provided: 510-724-70 |
| Quantity | 206 |
| Distribution | Worldwide distribution - U.S Nationwide distribution including in the states of CA, FL, NJ, IL, NE, TN, IN, TX, LA, WA, MI, MA, MN, OH, HI, MO, PA, CT, UT, SD, NY, AL, PR, OR, MD, KY, NC, AZ, SC, VA, WI, GA, IA, RI, AK, OK, KS, ND, MS, ME, NM, AR and the countries of ES, AE, DE, FR, KR, CN, TW, JP, |
| FDA record | Open on fda.gov |