Recall Z-0076-2022 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp.

Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

DateOct 7, 2021 (initiated Sep 8, 2021)
StatusOpen, Classified
ReasonThe Helium Indicator on the Cardiosave display may over represent the amount of helium in unit in the presence of Electromagnetic Interference.
Root causeDevice Design
ActionOn Wednesday, September 8, 2021, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that all lots of Maquet Cardiosave Hybrid and Maquet Cardiosave Rescue experience the following issue: Helium Indicator Inaccuracy in the Presence of Continuous EMC Disturbances. Customers were provided with actions to take to limit the possibility of loss of therapy due to an over-reporting of remaining helium on the Helium Indicator. Cardiosave Hybrid users should view the mechanical gauge on the Cardiosave Top Panel to assess remaining helium capaci
Quantity5054 total (US); 5033 total (OUS)
DistributionWorldwide distribution - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181122CARDIOSAVE Intra-Aortic Balloon Pump
Datascope Corp., May 31, 2018