Recall Z-0151-2022 ยท N/A

Instrumentation Laboratory

HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma on IL Coagulation Systems, (ACLTOP Family, ACL TOP Family 50 Seriest and ACL Elite/Elite Pro) Part Number: 0020302601

DateAug 25, 2021
StatusOpen, Classified
ReasonLabeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days
Root causeDevice Design
ActionWerfen issued Urgent Medical Device Correction letter via email via email or registered mail on 9/7/21. Letter states reason for recall, health risk and action to take: NOTE: Please refer to the separate "URGENT Medical Device Removal" notice (dated August 25, 2021) for Hemosll Liquid Anti-Xa, Part No. 0020302600, which removes claims for this assay from the ACL Elite and ACL Elite Pro instruments. " Mandatory Customer Actions At this time, the on-board instrument stability claim has been reduced from 7 days to 5 days for all in-date and future product lots of Hemosll Liquid Anti-Xa (Part Nos
Quantity40,224
DistributionUS Nationwide Foreign: Canada Armenia Austria Australia Belgium Brunei Brazil Belarus Switzerland China Colombia Czech Republic Germany Denmark Spain Finland France United Kingdom Gibraltar Gree
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.