Recall Z-0121-2022 ยท FLE Sterilizer, Steam

Fort Defiance Industries, LLC

P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641

DateAug 19, 2021
StatusOpen, Classified
ReasonThe rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
Root causeDevice Design
ActionA Recall notification letter titled, "MEDICAL DEVICE ADVISORY NOTICE" was sent to consignees on 08/19/2021 via a email. The notice provides instructions regarding continued use of the device until FDI provides a protective cover for the rear heater plate and cover assembly. It mentions, "THE OPERATOR SHOULD NOT TOUCH THE REAR HEATER PLATE AND COVER ASSEMBLY OF THE DEVICE WHILE THE DEVICE IS IN USE OR POWERED ON. IF ACCESS TO THE REAR OF THE DEVICE IS REQUIRED, PLEASE TURN THE POWER OF THE UNIT OFF USING THE FRONT PANEL SWITCH OR DISCONNECT INCOMING POWER TO THE DEVICE. DO NOT ATTEMPT TO MODIFY
Quantity380 devices
DistributionDistributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141009FORT DEFIANCE AUTOMATED STEAM STERILIZER
Fort Defiance Industries, Inc., Dec 16, 2014