Recall Z-0070-2022 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound Inc

Philips Ultrasound, Inc. Sparq, Diagnostic Ultrasound System, REF 795090, 100-240V -, 50/60 Hz, 500 VA, CE2797, Rx Only,

DateAug 19, 2021
StatusTerminated
ReasonBattery system data issue with the ultrasound system can intermittently cause a system shutdown, regardless of actual battery state or application of alternating current power.
Root causeSoftware design
ActionOn August 24, 2021, Philips issued "Urgent Medical device Correction" notification to affected consignees. The customer communication was distributed via: USPS Certified Mail In addition to informing consignees about the recall, Philips requested consignees take the following actions: Instructions for how to determine the software version of your Ultrasound system: 1. Power up the system and allow it to complete the boot sequence, 2. Press the Wrench icon on the upper right side of the control panel, 3. Press the Service button on the bottom left corner, 4. Under System Management, Press Sys
Quantity34 units
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, MA, MD, MI, MS, NC, NY, OH, VA, WA and the countries of Australia, Portugal, Netherlands, Canada, Netherlands, Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201012EPIQ Diagnostic Ultrasound System Series, Affiniti Diagnostic Ultrasound System Series, CX50 Diagnostic Ultrasound Syste