Recall Z-2582-2021 ยท ITX Transducer, Ultrasonic, Diagnostic

North Coast Medical Inc

Norco Ultrasound Gel, Item/Description: NC70479/Norco Ultrasound Gel, 250 mL; NC70479/North Coast Ultrasound Gel, 250 mL; NC70479-5/Norco Ultrasound Gel, 5 liters; NC70479-5/North Coast Ultrasound Gel, 5 liters; NC70479-5C/Norco Ultrasound Gel, 5 liters (4); NC70479-5C/North Coast Ultrasound Gel, 5 liters (4); NC70479C/Norco Ultrasound Gel, 250 mL (24); NC70479C/North Coast Ultrasound Gel, 250 mL

DateAug 27, 2021
StatusOpen, Classified
ReasonUltrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA published a Letter to Health Care Providers to further address this issue: https://www.fda.gov/medical-devices/letters-health-care-providers/stop-using-all-eco-med-ultrasound-gels-and-lotions-due-risk-bacterial-contamination-letter-health
Root causeProcess control
ActionOn 08/17/2021, an online posting went live informing customers to discontinue use of affected products. On 08/27/21, recall notices were mailed to customers. Customers were asked to do the following: 1) Immediately examine your inventory and discontinue use, quarantine, and stop distribution of all affected gels and lotions manufactured by Eco-Med Pharmaceutical. 2) If affected product has been further distributed, notify customers, and forward the recall notification down to the user level. If you have any questions concerning this communication, please don t hesitate to contact NCMRecall@n
Quantity576
DistributionWorldwide distribution. US nationwide, Canada, Israel, South Africa, Australia, South Korea, France, Singapore, Grand Cayman, and Switzerland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955246ECOGEL 200 ULTRASOUND GEL
Eco-Med Pharmaceuticals, Inc., Feb 21, 1996