Recall Z-2577-2021 · NRY Catheter, Thrombus Retriever

IMPERATIVE CARE INC

Imperative Care, ZOOM 71 Reperfusion Catheter, REF: ICRC071137, TRX Tip, 137 cm, 6F (.083 in/2.11 mm), .071 in/1.8 mm), STERILE.EO, Rx Only, UDI: (01) 00812212030191. For neurovascular procedures.

DateSep 29, 2021 (initiated Aug 13, 2021)
StatusCompleted
ReasonThere is a potential for distal end of catheters to fracture and become detached.
Root causeUnder Investigation by firm
ActionOn August 18, 2021 , Imperative Care issued a "Urgent: Voluntary Medical Device Recall" Notification to all affected healthcare facilities. In addition to informing consignees about the recalled product, Imperative Care ask consignees to take the following actions: 1. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. 2. Immediately review your inventory for the specific lot numbers listed within the recall notification . 3. Remove and quarantine all unused affected products in your inventory. 4. Return
Quantity3118 units
DistributionUS nationwide distribution.
FDA recordOpen on fda.gov