Recall Z-2581-2021 ยท KPT Calibrator, Dose, Radionuclide

Biodex Medical Systems, Inc.

AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332

DateAug 16, 2021
StatusTerminated
ReasonWhen deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration
Root causeSoftware design
ActionBiodex issued Urgent Medical Deice Recall Letter on 8/16/21 via email. Letter states reason for recall, health risk and action to take: Review the dial settings associated with custom isotope and confirm they have not been changed. Compare the measured activity with a dose calibrator from another manufacturer The error can be immediately corrected by restarting the system. We recommend that you promptly download new software. The error is corrected in Software Revision 2.0.10. The corrected software upgrade is available as a download from the Biodex website at the following link Atomlab 500 Do
Quantity429 units
DistributionNationwide Foreign: Austria Autralia Brazil CANADA Chile Colombia Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt France Hong Kong HUNGARY Iceland India Iraq Ireland ISRAEL Isreal Italy Japan JORDAN Kuwait Lebanon Libya Malaysia MALTA Mexico Morocco Netherlands New Zealand Oma
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090296ATOMLAB 400 AND ATOMLAB 500 DOSE CALIBRATORS, MODELS 086-330, -331, -332, -335 AND -336
Biodex Medical Systems, Inc., Feb 19, 2009