Recall Z-2612-2021 ยท MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate

Corin Ltd

Corin Unity Knee Patella, Size 3 / 8.5mm, Centered Dome 3 PEG-UHMWPE, REF 112.018.06, Sterile, a component of the Unity Total Knee System.

DateJul 28, 2021
StatusTerminated
ReasonThe packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.
Root causeStorage
ActionOn 7/28&29/2021, the firm sent emails entitled "IMPORTANT-containment request FA COUK 2021 013" which notified the consignee they have been identified as receiving one or more of the medical devices listed in the attached Excel file due to them potentially having physical and/or water damage. They were to quarantine the parts until further instructions were issued. On 9/7/2021, Corin USA issued recall letters dated 9/3/2021 via email on 9/7/2021 requesting the devices be quarantined and returned.
Quantity1 unit
DistributionUS Nationwide distribution in the states of AZ, CA, FL, IL, OK, PA, TX, UT, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.