Recall Z-2446-2021 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

syngo.CT VA20A, VA30A and VA40A in: syngo.via RT Image Suite / Sim&Go SOMATOM go.Now SOMATOM go.Up SOMATOM go.All SOMATOM go.Top SOMATOM go.Sim SOMATOM go.Open Pro and SOMATOM X.cite Material Numbers: 11061620 11061628 11061630 11061640 11061660 11061670

DateSep 10, 2021 (initiated Aug 16, 2021)
StatusOpen, Classified
ReasonPotential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Root causeSoftware design
ActionSiemens has initiated three (3) Customer Safety Advisory Notices (CSANs) to affected customers via SY052/21/S, CT042/21/S and MR019/21/S by email or hand delivery by the Siemens Customer Service Engineer on 8/16/21. Temporary workarounds are available for some of the software versions and are described in the CSANs. Additionally, a solution to eliminate the root cause of this problem is being distributed via SY051/21/S (for SY and CT) and MR021/21/S for MR to all affected customers free of charge. The service organization will contact you to arrange a date to perform this corrective action f
Quantity44 systems
DistributionUS Nationwide distribution.