ClearanceBrief
Recall Z-2567-2021 ยท EOQ Bronchoscope (Flexible Or Rigid)

Olympus Corporation of the Americas

BF-1T60 OES Bronchofiberscope, Model No. BF-1T60

DateAug 16, 2021
StatusTerminated
ReasonThe firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
Root causeReprocessing Controls
ActionOn August __, 2021, the firm distributed Urgent Medical Device Corrective Action letters to customers. The applicable addenda to the reprocessing instructions were attached to the letters. Action steps to be taken by the end user: 1. Inspect your inventory for the referenced devices and identify any device with the model names specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Carefully read the content of this Medical Device Correction Action as well as the attached Addendum . The addendum provides updated information o
Quantity350
DistributionDomestic distribution nationwide. Product also distributed globally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023984MODIFICATION TO OLYMPUS BRONCHOSCOPES BF-40 SERIES, BF-240 SERIES, AND BF-160 SERIES
The Olympus Optical Co., Jan 31, 2003