Recall Z-0032-2022 ยท MMI Immunoassay Method, Troponin Subunit

Beckman Coulter Inc.

Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK

DateAug 9, 2021
StatusOpen, Classified
ReasonThrough customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.
Root causeUnknown/Undetermined by firm
ActionDuring the week of 08/09/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via mail or email informing customer of possible carryover with in-use, open (punctured) hsTnI reagent packs, and the impact of carryover on patient samples that are tested from the same reagent pack as a sample with a high cardiac troponin (cTnI) concentration >270,000 pg/mL (ng/L), or the carryover may also impact a different reagent pack. Clinical carryover can affect the results of all subsequent samples that are tested from the affected pack. Customer are instructed to: - If an hsTnI resul
Quantity293,640 kits
DistributionU.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzeg
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172783Access hs Tnl
Beckman Coulter, Inc., Jun 12, 2018
K172787Access hsTnl
Beckman Coulter, Inc., Jun 14, 2018