Recall Z-2587-2021 ยท MRN Apparatus, Nitric Oxide Delivery

Vero Biotech, LLC

GENOSYL DS (Delivery System) console, for use as a vasodilator.

DateSep 9, 2021
StatusCompleted
ReasonManufacturing defect involving reversal of the inlet and outlet gas lines, resulting in delivery of nitrogen dioxide (NO2) instead of nitric oxide (NO).
Root causeProcess control
ActionOn 09/09/2021, Vero Biotech notified customers via certified mail with letter titled, "URGENT: MEDICAL DEVICE RECALL GENOSYL DS LIMITED MANUFACTURING DEFECT." The letter notified the customer of the issue and there were no actions to be taken as the three affected devices were removed from the field prior and replaced.
Quantity120 consoles (3 consoles affected)
DistributionDistributed within US to Texas and Illinois.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.