ClearanceBrief
Recall Z-2600-2021 ยท FSA Lift, Patient, Non-Ac-Powered

Hill-Rom, Inc.

LikoStretch 1900 Model Number: 3156051 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.

DateAug 11, 2021
StatusOpen, Classified
ReasonStretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.
Root causeUnder Investigation by firm
ActionHillrom issued URGENT Medical Device Correction letter dated August 11, 2021 by third party vendor via UPS mail/track the delivery to consignees. Letter states reason for recall, health risk and action to take: Hillrom is initiating this Medical Device Correction to all customers and will be correcting devices with an updated to the leveller. Actions being taken by Customer: Please share this Medical Device Correction with all potential users in your organization. Please ensure inspection of the device prior to use as stated in the device Instruction for Use (IFU). Complete the attached respo
Quantity9 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, DE, IL, ME, MI, MO, MS, NJ, OH, PA, VA, WA and the countries of Argentina, Australia, Austria, Bolivia, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, G
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.