ClearanceBrief
Recall Z-2476-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Philips North America Llc

The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: "Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. "Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery p

DateAug 13, 2021
StatusTerminated
ReasonEarly life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Root causeDevice Design
ActionPhilips notified accounts by Urgent Medical Device Recall letter via certified mailing United States Postal Service (USPS) on 8/13/21. Letter states reason for recall, health risk and action to take: Please place this Medical Device Recall letter with the documentation of the system until this correction has been implemented in your system. Additionally, please return the attached reply form to Phhilps to confirm that the users of the system have reviewed and understood this Medical Device Recall. Philips will correct the affected systems by replacing the ADUs. A Philips service representat
Quantity1 unit
DistributionUS Nationwide distribution in the states of ID, MA, MI, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162148Allura Xper R9
Philips Medical Systems Nethrlands BV, Nov 23, 2016