Recall Z-0389-2022 ยท JAD System, Therapeutic, X-Ray

Xstrahl Limited

Treatment applicator for Gulmay Medical 150 or Xstrahl 150 superficial X-ray therapy system

DateAug 27, 2021
StatusTerminated
ReasonBase plate may detach from the main body of the treatment applicator.
Root causeDevice Design
ActionThe firm notified their consignees by email on 08/27/2021. The notice requested the following actions be taken: Xstrahl Limited are issuing the following inspection advice for all X150 specifical system users. 1. Prior to any clinical exposure the operator of the system should inspect the treatment applicator prior to inserting it into the treatment head of the X150. 2. As part of the daily assurance process the treatment applicators should be inspected. 3. If a treatment applicator is dropped it should be inspected by medical physics to ensure it is not damaged. 4. If the base plate is mis
Quantity120 applicators
DistributionUS and ALGERIA, AUSTRALIA, AUSTRIA, BULGARIA, CANADA, COSTA RICA, CROATIA, CYPRUS, DENMARK, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KAZAKHSTAN, LATVIA, MALTA, MEXICO, NEW ZEALAND, NORWAY, PANAMA, ROMANIA, RUSSIA, SAUDI ARABIA, SPAIN, SWEDEN, SWITZERLAND, UK, UKRAINE