Recall Z-2391-2021 ยท ITX Transducer, Ultrasonic, Diagnostic
Athena Medical Products Inc
LiquaSonic Ultrasound Gel 250mL, Model No. 001205
| Date | Sep 9, 2021 (initiated Aug 20, 2021) |
| Status | Terminated |
| Reason | Ultrasound gels and lotions manufactured by Eco-Med Pharmaceutical, Inc. may contain bacterial contamination. On August 18, 2021, FDA published a Letter to Health Care Providers to further address this issue: https://www.fda.gov/medical-devices/letters-health-care-providers/stop-using-all-eco-med-ultrasound-gels-and-lotions-due-risk-bacterial-contamination-letter-health. |
| Root cause | Process control |
| Action | On August 20, 2021, the firm issued an Urgent Medical Device Recall letter to its only customer, who is a distributor. The letter explained the product issue and asked that the distributor identify their customers and notify them at once of the product recall. ALL lots of product should be returned or destroyed. Athena will no longer purchase or distribute products manufactured by Eco-Med. If you or your customers have any questions you can call the Recall Coordinator at 888-216-7307 or email athena8649@stericycle.com. |
| Quantity | 12,240 cases of 12 |
| Distribution | Distributed to one distributor within the US. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K955246 | ECOGEL 200 ULTRASOUND GEL Eco-Med Pharmaceuticals, Inc., Feb 21, 1996 |