ClearanceBrief
Recall Z-2593-2021 ยท KPS System, Tomography, Computed, Emission

Shanghai United Imaging Healthcare Co., Ltd.

uMI 550 System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

DateAug 2, 2021
StatusOpen, Classified
ReasonThe EXAM acquisition workflow could occasionally freeze after canceling the PET scan.
Root causeDevice Design
ActionOn 08/02/2021, the Firm emailed to its customer an "Advisory Notification" informing them of a potential safety-relevant issue in which the EXAM acquisition workflow could occasionally freeze after canceling the PET scan. The issue and the sequence of events leading to this occurrence are as follow: During a myocardial perfusion scan procedure, the follow steps could lead to the EXAM acquisition workflow to occasionally freeze- 1. Selecting the myocardial perfusion scan procedure and entering EXAM. 2. Finishing the CT scout and helical/cine scan. 3. Starting the PET scan. 4. Canceling the cur
Quantity23 systems
DistributionU.S. Nationwide distribution in the states of CA, IL, IN, TX, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182237uMI 550 PET/CT System
Shanghai United Imaging Healthcare Co., Ltd., Oct 12, 2018
K193241Emission Computed Tomography System
Shanghai United Imaging Healthcare Co., Ltd., Jan 26, 2020