Recall Z-2574-2021 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Medline Industries Inc

NAMIC Manifolds. (1) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038301, UPN H965640383011, 3 Valve Manifold, Off Star Handles, Ports on Right, Rx only, Sterile; (2) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038303, UPN H965640383031, 3 Valve Manifold, On Star Handles, Ports on Right, Rx only, Sterile; (3) NAMIC Angiographic MANIFOLD High Pressure, REF 70037200, UPN H96570

DateAug 6, 2021
StatusTerminated
ReasonPotential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions.
Root causeProcess design
ActionThe recalling firm issued letters dated and mailed on 8/6/2021 notifying consignees of the issue and requesting they immediately check their stock for the affected product and destroy it.
Quantity816,420 manifolds and stopcocks
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Australia, Japan, Latvia, Lebanon, Malaysia, Netherlands, Pakistan, People's Republic of China, Panama, Singapore, South Africa, South Korea, Thailand, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K782095STOPCOCKS, MANIFOLDS
Namic, Jan 23, 1979
K842829HIGH PRESSURE, 1 & 3 WAY STERILE DISPOS
North American Instrument Corp., Oct 18, 1984