Recall Z-0011-2022 ยท DPS Electrocardiograph

Welch Allyn Inc Mortara

ELI, Burdick and McKesson brand 280 Resting Electrocardiographs (ELI 280, MLBUR280 and BUR280), Model Numbers: ELI280-DDB-ADAAX, ELI280-BDB-ACAAX, ELI280-CAA-AAFAD, ELI280-BDB-BDFAX, ELI280-BCB-AAAAX, ELI280-DDB-ACAAX, ELI280-DBA-BAFAX, ELI280-DCB-AAAAX, ELI280-DBA-AAFBD, ELI280-AAA-AAEBX, ELI280-BBA-ADFBD, ELI280-DDB-ACFAX, MLBUR280-81X, ELI280-CAA-ACEBX, ELI280-AAA-AAHBX, ELI280-DDB-BD

DateSep 3, 2021
StatusOpen, Classified
ReasonThe devices malfunction under specific operator workflows.
Root causeSoftware design
ActionThe firm notified its consignees on 9/3/2021 by letter via U PS or via USPS first class for those with P.O. Box mailing addresses. The letter notified customers of the issue, provided clinical mitigations and advised them not to utilize specific workflows until their software is updated.
Quantity14,630 units
DistributionDistribution was nationwide, including Puerto Rico. Foreign distribution was made to Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Czechia, Denmark, Dominican Republic,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122073ELI 280 ELECTROCARDIOGRAPH
Mortara Instrument, Inc., Aug 2, 2012