Recall Z-2472-2021 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories

Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use. Model number LN 03R70-01

DateJul 26, 2021
StatusOpen, Classified
ReasonDue to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on the Reagent Status screen with a status of "OK". In this scenario, the expired reagent will remain in the carousel for sample processing. 2) Performing monthly maintenance procedure 5701 Clean ICT Drain Tip incorrectly may potentially cause issues such as damaged connector or leaking connection. 3) A sample exception with message code 150 "Unable to process tes
Root causeSoftware design
ActionOn 07/26/2021, the firm sent an "Product Correction Urgent-Immediate Action Required" Notification Letter via FedEx priority overnight express informing customers that the Recalling Firm identified three potential performance issues with software version 3.2.3 and earlier and that the Recalling Firm is releasing software version 3.3.0 to correct the following issues: 1. There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on the Reagent Status screen with a status of OK . In this scenario, the expired reagent will rem
Quantity4,833 systems
DistributionWorldwide distribution. US nationwide including Puerto Rico; Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.