ClearanceBrief
Recall Z-2525-2021 ยท MJC Catheter, Urological (Antimicrobial) And Accessories

C.R. Bard Inc

Catalog A303316A, SureStep" Foley Tray, Bardex" I.C., Complete Care", Urine Meter, StatLock" Stabilization Device

DateSep 23, 2021 (initiated Aug 5, 2021)
StatusOpen, Classified
ReasonThere is potential for packaging defects that may impact the sterile barrier.
Root causePackaging change control
ActionCustomer letters were sent "URGENT MEDICAL DEVICE RECALL" via FedEx and email to affected consignees on August 13, 2021. Customers are asked to please Take the Following Actions: 1. Note that there may be a supply disruption in Foley catheter trays from all sources due to the current product shortage driven by this issue and other issues with third party Foley manufacturers. 2. Please check all inventory locations within your institution for the specific catalog and lot number combinations listed in Attachment 1 of this recall notice. 3. Share this recall notice with any users of the produc
Quantity11,600 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040658BARDEX I.C. LATEX FOLEY CATHETER
C.R. Bard, Inc., Sep 29, 2004