ClearanceBrief
Recall Z-2412-2021 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

Ingenia 1.5T Evolution-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Model: 781315

DateAug 11, 2021
StatusCompleted
ReasonWhen using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
Root causeDevice Design
ActionPhilips notified U.S. customers an Urgent Medical Device Correction letter on 8/11/21. Letter states reason for recall, health risk and action to take: Philips advises that customers continue to utilize the primary motorized function for the MR system since the identified issue with the HDU does not occur in the motorized mode. If you utilize the manual function and it is not functioning properly, then you must perform table movement using the motorized function. A Philips Field Service Engineer (FSE) will perform a HDU during the next scheduled preventive maintenance visit. If the issue is de
Quantity42 units
DistributionWorldwide distribution - US Nationwide distribution and the countries of Albania, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162931Ingenia 1.5T CX and Ingenia 3.0T CX R5.3
Philips Medical Systems Nederland B.V., Jan 6, 2017