Recall Z-2482-2021 · FLL Continuous Measurement Thermometer

Welch Allyn Inc

Braun ThermoScan PRO 6000 Ear Thermometer-indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment. Model: 6000

DateJul 20, 2021
StatusOpen, Classified
ReasonIf the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the user or the ear canal of the patient
Root causeLabeling design
ActionWelch Allen (a Hill Rom company) issued Urgent Medical Device Correction Letter (FA-2021-05-001-SKF-004) on 7/30/21 to: Chief Executive; Facility Administrator; Facility Engineer; Vigilance Manager; Biomedical Engineering; Medical Device Liaison Officer; Distributor mailed (with signature service) to all first-level consignees by Sedgwick (third party recall company). The letter states reason for the Field Action, risk mitigations, and Actions to be taken by the User: 1. Please share this communication with all potential users in your organization and instruct them to follow the provided Cle
Quantity1,036,740 units
DistributionWorldwide distribution - US Nationwide distribution and the countries of Afghanistan, Andorra, Australia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Canada, Chile, China, Colombia, Congo, Costa Rica, C¿te D'ivoire (Iv
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152748Braun Thermoscan PRO 6000 Ear Thermometer
Kaz USA, Inc (A Subsidiary of Helen of Troy, Inc), Mar 1, 2016