Recall Z-0002-2022 ยท FMI Needle, Hypodermic, Single Lumen

Bard Peripheral Vascular Inc

REF TIN3015, 15G x 79 mm Bone Marrow Aspiration/Intraosseous Infusion Needle, For Single Use Only, Sterile EO, UDI: 10885403043796

DateJul 27, 2021
StatusOpen, Classified
ReasonPotential for needle protector to become fully dislodge exposing infusion needle which could result in needle stick.
Root causePackaging change control
ActionOn August 4, 2021, Bard Peripheral Vascular, Inc. issued an "Urgent Medical device Correction" notice to affected customers via Fed Ex. In addition to informing consignees about the recall product, BD ask consignees to take the following actions: 1. Please check all inventory locations within your institution for the product listed in Attachment A and immediately destroy all affected product remaining in your possession. The product should have the sterile barrier broken and destroyed per your local facility s destruction process. 2. Share this notice with any users of the product within yo
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WV and the countries of Netherlands, United Arab Emirates, Australia, Taiwan,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930217BAXTER DISPOSABLE INTRAOSSEOUS INFUSION NEEDLE
Baxter Healthcare Corp, Feb 28, 1994