ClearanceBrief
Recall Z-2346-2021 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

Achieva 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781277 (OUS)

DateJul 30, 2021
StatusTerminated
ReasonPressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury
Root causeNonconforming Material/Component
ActionPhilips North America issued U.S. customers letters via certified mail using a 3rd Party service. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall , health risk and action to take: The Philips Ingenia 1.5T, Ingenia 1.5T Evolution, Ingenia 3.0T, and Ingenia Elition X MR systems can continue to fulfill clinical use as normal. However, in case of technical problems related to off cooling of the magnet, magnet pressure or Helium boil-off, please stop use and contact your local Philips service representative immediately a
Quantity1 unit
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Neth
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110151ACHIEVA R4 1.5T AND ACHIEVA R4 3.0T
Philips Health Care, Mar 22, 2011