Recall Z-2434-2021 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

Proteus 235- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation

DateSep 8, 2021 (initiated Jun 14, 2021)
StatusOpen, Classified
ReasonUsers may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment
Root causeDevice Design
ActionIBA issued an Urgent Medical Device Correction notification users of the device, starting June 14, 2021. Letter states reason for recall, health risk and action to take: IBA would like to remind Proteus235 users to verify with X-ray images or other means, after the display of the popup message, that the patient positions (represented by the couch position, mainly the 3 translations and 3 rotations parameters) recorded in the OIS are correct Labelling change IBA will update the user manuals to: - Describe under which conditions the popup message appears and the subsequent risks, especially that
Quantity27 worldwide, 11 in U.S.
DistributionWorldwide distribution - US Nationwide distribution in the states of FL, OK, VA, IL, NJ, TN, WA, LA, TX, MI and the countries of Germany, Italy, Russia, Sweden, India, Japan, France, England, Spain.